Panel Introduction of SAEU3H Chassis and Installation of various AE boards.

Brief:SAEU3H adopts plug-and-plug structure design and can be equipped with different number of AE boards on the chassis to achieve different number of channels according to different needs.


1. Panel introduction of SAEU3H chassis

SAEU3H adopts plug-and-plug structure design and can be equipped with different number of AE boards on the chassis to achieve different number of channels according to different needs.

    Network communication boardoptional and detachable. For data communication between the unit and analysis software.

    EX-parameter 4-channel AE boardoptional and detachable. For AE parameter and external parameters collection.

    Standard 4-channel AE boarddetachable. For AE parameters collection.

    Channel indicator: There are 8 indicators on each AE board, and 2 indicators for each channel from top to bottom; The green light on the left is the signal threshold indicator, which will shine when the threshold is triggered by the acoustic emission signal; The yellow light on the right side is preamplifier status indicator. After the AE board is correctly identified, the preamplifier status indicator will light up.

    Female RJ45 connector

    16-pin LEMO connector

    C5 LEMO connector

    Chassis status indicators

    NumberIndicates the number of AE boards ahead of this chassis when the chassis is cascaded.

    Cover board: detachable. Extra AE board could be inserted when the channel is not enough.

 

    Power status indicator

    Switch

    Power supply connector: AC 100-240V/50-60Hz.

    Ground connector

    UP HDMI connector: For connecting with previous chassis when the chassis is cascaded.

    DN HDMI connector: For connecting with latter chassis when the chassis is cascaded.

    USB connector: For data communication between the unit and analysis software.

PDF File of Panel Introduction Sheet




 

2. Installation of various AE boards

 
Standard 4-channel AE board: Each AE board contains 4 channels; Users can select the number of acquisition boards according to different needs.

Network communication board: The SAEU3H chassis has two different communication modes: USB3.0 and Ethernet. The default configuration is USB3.0 communication. If you choose Ethernet for data communication, you need to select the network communication board for communication. Please note: When choosing to use the network communication board, the network communication board needs to occupy one board slot, and only occupy the leftmost board slot.

EX-parameters 4-channel AE board: has one more external parameter acquisition channel than the standard 4-channel AE board, and the interface is a 16-pin LEMO port, which can acquire the external ±10V voltage signal and acoustic emission signal synchronously. If this board is selected, it needs to occupy the board slot before the standard 4-channel AE board, and it needs to be used with EX-PARA box.



 

Contact

Hotline:+86-20-32290092Cel:+86 19128609186WeChat/Whatsapp:+86 19128609186Email: sales2@ae-ndt.comSkype:jessicajia

9:00-18:00 (UTC+8)

Follow Us

YouTube

Linkedin

Facebook

© Copyright 2024 QAWRUMS Ltd. All rights reserved. Privacy Policy